Image quality assessment of coronary artery segments using ultra-high resolution dual source photon-counting detector
Judith van der Bie1, Simran P Sharma2, Marcel van Straten1
1Department of Radiology & Nuclear Medicine, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands.
Purpose:
The study is intended to assess the image quality of ultra-high resolution (UHR) coronary computed tomography angiography (CCTA) performed on dual source photon-counting detector CT (PCD-CT).
Method:
Consecutive patients, who underwent clinically indicated CCTA on PCD-CT (UHR 120x 0.2 mm collimation), were included. CCTA images were reconstructed at 0.2 mm slice thickness with Bv40, Bv44, Bv48 and Bv56 kernels and quantum iterative reconstruction level 4. Contrast-to-noise (CNR) and signal-to-noise ratios (SNR) were quantified from contrast-enhanced blood and subcutaneous adipose tissue. All reconstructions were scored per coronary segment (18-segment model) for presence, image quality, motion artefacts, stack artefacts, plaque presence and composition, and stenosis degree. Image quality was scored by two independent observers.
Results:
Sixty patients were included (median age 62 [25th - 75th percentile: 53-67] years, 45% male, median calcium score 62 [0-217]). The mean heart rate during scanning was 71 ± 11 bpm. Median CTDIvol was 19 [16-22]mGy and median DLP 243 [198-327]mGy.cm. The SNR was 9.3 ± 2.3 and the CNR was 11.7 ± 2.6. Of the potential 1080 coronary segments (60 patients x 18 segments), 255/256 (reader1/reader2) segments could not be assessed for being absent or non-evaluable due to size. Both readers scored 85% of the segments as excellent or very good (Intraclass Correlation Coefficient: 0.88 (95% CI: 0.87-0.90). Motion artefacts were present in 45(5%) segments, stack artefacts in 60(7%) segments and metal artefacts in 9(1%) segments.
Conclusion:
UHR dual-source PCD-CT CCTA provides excellent or very good image quality in 85% of coronary segments at relatively high heart rates at moderate radiation dose with only limited stack artefacts.
More Related Videos
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
06:18Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...