Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
Multi-Method Investigation of Blood Damage Induced By Blood Pumps in Different Clinical Support Modes
Yuan Li1, Xinyu Liu, Anqiang Sun
1From the Key Laboratory of Biomechanics and Mechanobiology (Beihang University), Ministry of Education, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, Beijing, China.
Abstract:
To investigate the effects of blood pumps operated in different modes on nonphysiologic flow patterns, cell and protein function, and the risk of bleeding, thrombosis, and hemolysis, an extracorporeal blood pump (CentriMag) was operated in three clinical modalities including heart failure (HF), venous-venous (V-V) extracorporeal membrane oxygenation (ECMO), and venous-arterial (V-A) ECMO. Computational fluid dynamics (CFD) methods and coupled hemolysis models as well as recently developed bleeding and thrombosis models associated with changes in platelet and von Willebrand factor (vWF) function were used to predict hydraulic performance and hemocompatibility. The V-A ECMO mode had the highest flow losses and shear stress levels, the V-V ECMO mode was intermediate, and the HF mode was the lowest. Different nonphysiologic flow patterns altered cell/protein morphology and function. The V-A ECMO mode resulted in the highest levels of platelet activation, receptor shedding, vWF unfolding, and high molecular weight multimers vWF (HMWM-vWF) degradation, leading to the lowest platelet adhesion and the highest vWF binding capacity, intermediate in the V-V ECMO mode, and opposite in the HF mode. The V-A ECMO mode resulted in the highest risk of bleeding, thrombosis, and hemolysis, with the V-V ECMO mode intermediate and the HF mode lowest. These findings are supported by published experimental or clinical statistics. Further studies found that secondary blood flow passages resulted in the highest risk of blood damage. Nonphysiologic blood flow patterns were strongly associated with cell and protein function changing, blood damage, and complications.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: