Improved temporal performance and optical quantum efficiency of avalanche amorphous selenium for low dose medical
Corey Orlik1, Sébastien Léveillé2, Salman M Arnab2
1Stony Brook University, Health Sciences Center L4-120, Department of Radiology, Stony Brook, New York, United States.
Purpose:
Active matrix flat panel imagers (AMFPIs) with thin-film transistor arrays experience image quality degradation by electronic noise in low-dose radiography and fluoroscopy. One potential solution is to overcome electronic noise using avalanche gain in an amorphous selenium (a-Se) (HARP) photoconductor in indirect AMFPI. In this work, we aim to improve temporal performance of HARP using a novel composite hole blocking layer (HBL) structure and increase optical quantum efficiency (OQE) to CsI:Tl scintillators by tellurium (Te) doping.
Approach:
Two different HARP structures were fabricated: Composite HBL samples and Te-doped samples. Dark current and optical sensitivity measurements were performed on the composite HBL samples to evaluate avalanche gain and temporal performance. The OQE and temporal performance of the Te-doped samples were characterized by optical sensitivity measurements. A charge transport model was used to investigate the hole mobility and lifetime of the Te-doped samples in combination with time-of-flight measurements.
Results:
The composite HBL has excellent temporal performance, with ghosting below 3% at 10 mR equivalent exposure. Furthermore, the composite HBL samples have dark current and achieved an avalanche gain of 16. Te-doped samples increased OQE from 0.018 to 0.43 for 532 nm light. The addition of Te resulted in 2.1% first-frame lag, attributed to hole trapping within the layer.
Conclusions:
The composite HBL and Te-doping can be utilized to improve upon the limitations of previously developed indirect HARP imagers, showing excellent temporal performance and increased OQE, respectively.
More Related Videos
12:21Close-Space Sublimation-Deposited Ultra-Thin CdSeTe/CdTe Solar Cells for Enhanced Short-Circuit Current Density and Photoluminescence
Published on: March 6, 2020
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...