Development, Construction, and Evaluation of an Alternative Dosimetry Phantom for Computed Tomography

Lukmanda Evan Lubis1,2,3, Windi Dliya Najmah1, Yuni Muliyanti1

  • 1Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Indonesia.

PubMed
Summary

This study aimed to develop and test a new type of phantom for measuring CT radiation doses. The researchers mixed epoxy resin with iodine to create a material that mimics the radiological properties of PMMA, a standard phantom material. They tested 12 different mixtures and selected one with a 0.46% iodine concentration that closely matched PMMA’s electron density and atomic number. The new phantom was evaluated by comparing dose measurements at various CT tube voltages against the standard PMMA phantom. The results showed small dose discrepancies and statistical similarity between the two phantom types. The study concludes that the alternative phantom is a viable option for CT dosimetry and could be used in place of PMMA.