Related Experiment Video
Updated: Jun 8, 2026

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Three-dimensional intracardiac echocardiography for left atrial appendage sizing and percutaneous occlusion guidance
Domenico G Della Rocca1,2, Michele Magnocavallo1,3, Carola Gianni1
1Texas Cardiac Arrhythmia Institute, St. David's Medical Center, 3000 N Interstate Hwy 35 Suite 720, Austin, TX, USA.
Aims:
Left atrial appendage (LAA) imaging is critical during percutaneous occlusion procedures. 3D-intracardiac echocardiography (ICE) features direct visualization of LAA from multiple cross-sectional planes at a time. We aimed at reporting procedural success of 3D-ICE-guided LAA occlusion and the correlation between pre-procedural transoesophageal echocardiography (TEE) and intraprocedural 3D-ICE for LAA sizing.
Methods And Results:
Among 274 patients undergoing left atrial appendage occlusion (LAAO) with a Watchman FLX, periprocedural ICE guidance was achieved via a commercially available 2D-ICE catheter (220 patients) or a novel (NUVISION™) 3D-ICE one (54 patients). Primary endpoint was a composite of procedural success and LAA sealing at follow-up TEE. Secondary endpoint was a composite of periprocedural device recapture/resizing plus presence of leaks ≥ 3 mm at follow-up TEE. 3D-ICE measurements of maximum landing zone correlated highly with pre-procedural TEE reference values [Pearson's: 0.94; P < 0.001; bias: -0.06 (-2.39, 2.27)]. The agreement between 3D-ICE-based device selection and final device size was 96.3% vs. 79.1% with 2D-ICE (P = 0.005). The incidence of the primary endpoint was 98.1% with 3D-ICE and 97.3% with 2D-ICE (P = 0.99). 2D-ICE patients had a trend towards a higher incidence of periprocedural device recapture/redeployment (31.5% vs. 44.5%; P = 0.09). The secondary endpoint occurred in 31.5% of 3D-ICE patients vs. 45.9% of 2D-ICE ones (P = 0.065).
Conclusion:
Intracardiac echocardiography-guided LAAO showed a very high success, with no major adverse events. A very high level of agreement for LAA sizing was found between pre-procedural TEE and periprocedural 3D-ICE. 3D-ICE performed significantly better than 2D-ICE for FLX size selection and may provide better guidance during device deployment.
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization