Measuring the Behavioral Effects of Intraocular Scatter
Published on: February 18, 2021
Characterizing glare effects associated with diffractive optics in presbyopia-correcting intraocular lenses
Grzegorz Łabuz1, Ramin Khoramnia, Weijia Yan
1From the The David J Apple Center for Vision Research, Department of Ophthalmology, University Hospital Heidelberg, Heidelberg, Germany (Łabuz, Khoramnia, Yan, Auffarth, Naujokaitis, Tandogan); Netherlands Institute for Neuroscience, Royal Netherlands Academy of Arts and Sciences, Amsterdam, the Netherlands (van den Berg); Augenklinik Pallas, Olten, Switzerland (Tandogan).
Purpose:
To objectively quantify glare of intraocular lenses (IOLs) using a diffractive principle to extend the visual range and to identify models with increased susceptibility to inducing glare.
Setting:
David J Apple Laboratory, Heidelberg, Germany.
Design:
Laboratory investigation.
Methods:
Glare was assessed by means of a straylight parameter with a standard C-Quant intended for 7 degrees. In addition, 2 C-Quant modifications were used to test lower angles (ie, 2.5 degrees and 3.5 degrees). The following IOL models were assessed: PanOptix, AT Lisa Tri, Synergy, and Triumf, the latter 2 with chromatic aberration correction at distance. Straylight from trifocal IOLs was compared against a monofocal W-60R lens. The C-Quant test was performed through the studied IOLs by using additional optical components attached to their ocular.
Results:
Straylight (deg 2 sr -1 ) of the control was <1 at all tested angles, with the trifocal models showing comparable straylight at 7 degrees. At 3.5 degrees, Triumf's straylight increased to 15.5 ± 0.6, followed by Synergy (6.2 ± 1.1), PanOptix (4.1 ± 0.3), and AT Lisa Tri (2.0 ± 0.8). The chromatic aberration-correcting models demonstrated correspondingly higher straylight (Synergy: 18.8 ± 1.3; Triumf: 17.3 ± 0.5) at 2.5 degrees compared with PanOptix (4.3 ± 0.4), AT Lisa Tri (2.1 ± 0.1), and monofocal IOLs yielding minimal or no increase.
Conclusions:
Trifocal IOLs induced increased straylight, but it was limited to lower angles, which may cause difficulties detecting these effects using a standard clinical approach. The latest IOL designs featuring chromatic aberration correction at far focus seem more susceptible than the established trifocal IOLs to inducing a glare phenomenon.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
Related Concept Videos
Focusing of Light in the Eye
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Photoreceptors and Visual Pathways