Related Experiment Video
Updated: Jul 5, 2025

Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Influence of the Spread of the Electric Field on Speech Recognition in Cochlear Implant Users
Anna C Kopsch1, Torsten Rahne, Stefan K Plontke
1Department of Otorhinolaryngology, Head and Neck Surgery, Martin Luther University Halle-Wittenberg, University Medicine Halle, Halle (Saale), Germany.
Objective:
To investigate the correlation of word recognition with cochlear implant (CI) and spread of the electric field.
Study Design:
Prospective, noninterventional, experimental study.
Setting:
A tertiary referral center.
Patients:
Thirty-eight adult CI users with poor (n = 11), fair (n = 13), and good (n = 16) word recognition performance.
Main Outcome Measure:
Transimpedances were measured after 37 μs. Word recognition score was recorded at 65 dB SPL for German monosyllables in quiet. Transimpedance half widths were calculated as a marker for spread of the electric field.
Results:
Narrow and broad spread of the electric field, i.e., small and large half widths, were observed in all word recognition performance groups. Most of the transimpedance matrices showed a pattern of expansion along the diagonal toward the apical electrode contacts. Word recognition was not correlated with transimpedance half widths.
Conclusions:
The half width of the spread of the electric field showed no correlation with word recognition scores in our study population.