Related Experiment Video
Updated: Jul 5, 2025

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
Is the spread of excitation different between adults and children cochlear implants users?
Maria Valéria Schmidt Goffi-Gomez1, Francyelle Machado da Silva Corrêa2, Ana Tereza Magalhães2
1ENT Department, Cochlear Implant Group at Clinical Hospital of the Medical School at University of São Paulo, Av. Dr. Eneas de Carvalho Aguiar 255, São Paulo, Brazil. valeria.goffi@hc.fm.usp.br.
Purpose:
While cochlea is adult size at birth, etiologies and bone density may differ between children and adults. Differences in neural response thresholds (tNRT) and the spread of excitation (SOE) width may impact the use of artificial intelligence algorithms in speech processor fitting.
Aim:
To identify whether neural response telemetry threshold and spread of excitation width are similar in adults and children.
Methods:
Retrospective cross-sectional study approved by the Ethical Board. Intraoperative tNRT and SOE recordings of consecutive cochlear implant surgeries in adults and children implanted with Cochlear devices (Cochlear™, Australia) were selected. SOE was recorded on electrode 11 (or adjacent, corresponding to the medial region of the cochlea) through the standard forward-masking technique in Custom Sound EP software, which provides SOE width in millimeters. Statistical comparison between adults and children was performed using the Mann-Whitney test (p ≤ 0.05).
Results:
Of 1282 recordings of intraoperative evaluations, 414 measurements were selected from children and adults. Despite the tNRT being similar between adults and children, SOE width was significantly different, with lower values in children with perimodiolar arrays. Besides, it was observed that there is a difference in the electrode where the SOE function peak occurred, more frequently shifted to electrode 12 in adults implanted. In straight arrays, there was no difference in any of the parameters analyzed on electrode 11.
Conclusion:
Although eCAP thresholds are similar, SOE measurements differ between adults and children in perimodiolar electrodes.