Integrated-mode proton radiography with 2D lateral projections
Mikaël Simard1, Daniel G Robertson2, Ryan Fullarton1
1Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom.
Physics in Medicine and Biology
|January 19, 2024
Summary
A new proton radiography method uses 2D lateral cameras to improve image quality for proton therapy. This technique enhances water equivalent thickness (WET) mapping and range verification, aiding adaptive proton therapy.
Area of Science:
- Medical Physics
- Radiotherapy Technology
- Image Reconstruction
Background:
- Proton radiography generates water equivalent thickness (WET) maps for motion management and range verification in proton therapy.
- Current reconstruction methods using single cameras (distal or lateral range telescope) yield limited image quality.
Purpose of the Study:
- To introduce a novel reconstruction framework utilizing 2D information from two lateral view cameras.
- To enhance image quality in proton radiography compared to existing methods.
Main Methods:
- A new reconstruction framework employing two lateral view cameras was developed.
- Methods were compared using Geant4 Monte Carlo simulations with XCAT phantom and slanted edge phantoms.
- Experimental validation was performed using a plastic volumetric scintillator and various phantoms.
Main Results:
- The proposed method achieved a resolution of 0.33 lp mm⁻¹ on a simulated slanted edge phantom, an improvement over the distal method (0.24 lp mm⁻¹).
- Mean absolute error on WET maps for the XCAT phantom was reduced from 3.4 mm to 2.7 mm.
- Experimental data showed a 36% contrast increase with 2D lateral views compared to the range telescope method, with a mean absolute error of 0.4 mm for WET accuracy.
Conclusions:
- The proposed reconstruction framework with 2D lateral views significantly improves proton radiograph quality.
- This method is compatible with clinical beam settings and shows potential for adaptive proton therapy.
More Related Videos
Related Concept Videos
Positron Emission Tomography
4.2K
Positron emission tomography (PET) is a medical imaging technique involving radiopharmaceuticals — substances that emit short-lived radiation. Although the first PET scanner was introduced in 1961, it took 15 more years before radiopharmaceuticals were combined with the technique and revolutionized its potential.
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
4.2K
X-ray Imaging
5.5K
German physicist Wilhelm Röntgen (1845–1923) was experimenting with electrical current when he discovered that a mysterious and invisible "ray" would pass through his flesh but leave an outline of his bones on a screen coated with a metal compound. In 1895, Röntgen made the first durable record of the internal parts of a living human: an "X-ray" image (as it came to be called) of his wife’s hand. Scientists worldwide quickly began their own experiments with...
5.5K