The Repeatability of Axial Length Measurements Using a Scheimpflug-based System
Sara Sardari1,2, Mehdi Khabazkhoob3, Ebrahim Jafarzadehpur4
1Research and Technology Deputy, Tehran University of Medical Sciences, Tehran, Iran.
Journal of Ophthalmic & Vision Research
|January 22, 2024
Summary
The Pentacam AXL system shows high repeatability for axial length measurements. However, factors like male gender, older age, and nuclear cataracts can reduce measurement accuracy.
Area of Science:
- Ophthalmology
- Biometry
- Medical Devices
Background:
- Accurate axial length (AL) measurement is crucial for refractive error correction and intraocular lens power calculation.
- Scheimpflug-based devices offer advanced optical biometry, but their repeatability across diverse patient demographics requires thorough assessment.
Purpose of the Study:
- To evaluate the repeatability of axial length (AL) measurements using the Pentacam AXL, a Scheimpflug-based optical biometry system.
- To investigate the influence of patient age, sex, lens type, AL value, and cataract type on measurement repeatability.
Main Methods:
- Multistage cluster sampling was employed in Tehran, Iran, involving 897 eyes from 677 individuals (aged 20-91 years).
- Ocular biometry was performed thrice for each eye using the Pentacam AXL, with a 10-minute interval between measurements.
- Repeatability was assessed using the intraclass correlation coefficient (ICC) and repeatability coefficient (RC); tolerance index (TI) was used to compare repeatability across variables.
Main Results:
- The Pentacam AXL demonstrated high overall repeatability for AL measurements, with an ICC of 0.981 and an RC of 0.401.
- Repeatability was slightly better in females (ICC: 0.985) compared to males (ICC: 0.976).
- Measurement repeatability decreased with advancing age (ICC from 0.999 for <40 years to 0.973 for >60 years) and was reduced in eyes with nuclear cataracts and lower signal-to-noise ratios (SNR).
Conclusions:
- The Pentacam AXL system exhibits excellent repeatability for axial length measurements in a diverse population.
- Factors including male gender, older age, nuclear cataracts, and lower SNR are associated with decreased measurement repeatability.
- Optimizing measurement protocols and considering patient-specific factors can enhance the reliability of Pentacam AXL biometry.