The Impact of Dual and Triple Energy Window Scatter Correction on I-123 Postsurgical Thyroid SPECT/CT Imaging Using a
Konstantinos Michael1,2, Savvas Frangos3, Ioannis Iakovou4
1Department of Mechanical Engineering, Frederick University, Nicosia 1036, Cyprus.
Abstract:
I-123 is preferential over I-131 for diagnostic SPECT imaging after a thyroidectomy to determine the presence and size of residual thyroid tissue for radioiodine ablation. Scattering degrades the quality of I-123 SPECT images, primarily due to the penetration of high-energy photons into the main photopeak. The objective of this study was to quantitatively and qualitatively investigate the impact of two widely used window-based scatter correction techniques, the dual energy window (DEW) and triple energy window (TEW) techniques, in I-123 postsurgical SPECT/CT thyroid imaging using an anthropomorphic phantom with small sizes of remnants and anatomically correct surrounding structures. For this purpose, non-scatter-corrected, DEW and TEW scatter-corrected SPECT/CT acquisitions were performed for 0.5-10 mL remnants within a phantom, with 0.5-12.6 MBq administered activities within the remnants, and without and with background-to-remnant activity ratios of 5% and 10%. The decrease in photons, the noise and non-uniformity in the background region due to scatter correction were measured, as well as the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) from small remnants. The images were also visually evaluated by two experienced nuclear medicine physicians. Scatter correction decreased photons to a higher extent in larger regions than smaller regions. Larger remnants yielded higher SNR and CNR values, particularly at lower background activities. It was found from the quantitative analysis and the qualitative evaluation that TEW scatter correction performed better than DEW scatter correction, particularly at higher background activities, while no significant differences were reported at lower background activities. Scatter correction should be applied in I-123 postsurgical SPECT/CT imaging to improve the image contrast and detectability of small remnants within the background.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...