The Development of a Measuring System for Intraoral SpO2
Yuki Kashima1, Minako Onimaru1, Ryosuke Isogai2
1Department of Orthodontics, School of Dentistry, Showa University, 2-1-1 Kitasenzoku, Ota-ku, Tokyo 145-8515, Japan.
Abstract:
Blood oxygen saturation (SpO2) is an essential indicator of a patient's general condition. However, conventional measurement methods have some issues such as time delay and interference by ambient light. Improved measurement methods must be developed, and there are no reports on intraoral measurements of SpO2 using wearable devices. Therefore, we aimed to establish an intraoral SpO2 measurement method for the first time. Twelve healthy adults participated in this study. The following steps were taken: (1) to identify the optimal measurement location, mid-perfusion index (PI) values were measured at six places on the mucosa of the maxilla, (2) to validate the optimal measurement pressure, PI values were obtained at different pressures, and (3) using the proposed mouthpiece device, SpO2 values in the oral cavity and on the finger were analyzed during breath-holding. The highest PI values were observed in the palatal gingiva of the maxillary canine teeth, with high PI values at pressures ranging from 0.3 to 0.8 N. In addition, changes in SpO2 were detected approximately 7 s faster in the oral cavity than those on the finger, which is attributed to their proximity to the heart. This study demonstrates the advantage of the oral cavity for acquiring biological information using a novel device.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...