Related Experiment Video
Updated: Jul 5, 2025

Hydra, a Computer-Based Platform for Aiding Clinicians in Cardiovascular Analysis and Diagnosis
Published on: September 26, 2018
Clinical Validation of Computer-Aided Diagnosis Software for Preventing Retained Surgical Sponges
Ken Kurisaki1, Akihiko Soyama1, Shin Hamauzu2
1From the Department of Surgery, Nagasaki University Graduate School of Biomedical Sciences, Nagasaki City, Japan (Kurisaki, Soyama, Yamaguchi, Matsuguma, Eguchi).
Background:
We previously reported the successful development of a computer-aided diagnosis (CAD) system for preventing retained surgical sponges with deep learning using training data, including composite and simulated radiographs. In this study, we evaluated the efficacy of the CAD system in a clinical setting.
Study Design:
A total of 1,053 postoperative radiographs obtained from patients 20 years of age or older who underwent surgery were evaluated. We implemented a foreign object detection application software on the portable radiographic device used in the operating room to detect retained surgical sponges. The results of the CAD system diagnosis were prospectively collected.
Results:
Among the 1,053 images, the CAD system detected possible retained surgical items in 150 images. Specificity was 85.8%, which is similar to the data obtained during the development of the software.
Conclusions:
The validation of a CAD system using deep learning in a clinical setting showed similar efficacy as during the development of the system. These results suggest that the CAD system can contribute to the establishment of a more effective protocol than the current standard practice for preventing the retention of surgical items.