Single operator experience, learning curve, outcomes, and insights gained with conduction system pacing
William Marion1, John D Schanz2, Sonali Patel1
1Sidney Kimmel Medical College, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
Background:
Conduction system pacing (CSP) is increasingly utilized to prevent and correct dyssynchrony. Barriers to CSP adoption include limited training, methodologic variability, laboratory slot allocation, and few data on learning curves. We report learning curves/clinical outcomes from a single experienced electrophysiologist who was new to CSP, and share gained insights.
Methods:
Retrospective analysis of all patients who underwent attempted CSP implantation (2016-2023). Patient characteristics, ECGs, echocardiograms, fluoroscopy/procedure times, lead data were recorded at implant and follow-up.
Results:
CSP leads were implanted successfully in 167/191(87.4%) patients with a follow-up of 278 ± 378 days. His-bundle pacing (HBP = 59) and left-bundle-area pacing (LBAP = 108) had similar procedure/fluoroscopy times, QRS duration decreases, and ejection fraction improvements (all p > NS). Eight HBP lead revisions were required for high capture thresholds LBAP demonstrated lower pacing thresholds, higher lead impedances, and greater R-wave amplitudes at implant and follow-up. After 25 HBP cases, implant pacing thresholds, fluoroscopy, procedural times did not decrease. After 25 LBAP cases, there were significant decreases in all these parameters (p < 0.05). A separate analysis in LBAP patients with recorded Purkinje signals showed no differences in paced ECG characteristics between patients with pre- QRS Purkinje signals versus patients with Purkinje signals post-QRS onset.
Conclusions:
Experienced implanters who are new to CSP can achieve steady-state procedural/fluoroscopy times after a learning curve of 25 implants. LBAP showed lower capture thresholds and higher success rates. Adequate depth of lead deployment (as determined by published parameters) does not require Purkinje potential to be pre-QRS. Operators new to CSP.can forego HBP and directly implement LBAP.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrophysiology of Normal Cardiac Rhythm