Signal-correction errors in the EasyOne Pro LAB multiple-breath washout device significantly impact outcomes in
Marc-Alexander Oestreich1, Florian Wyler1, Christian Buess2
1Division of Paediatric Respiratory Medicine and Allergology, Department of Paediatrics, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
None:
Multiple-breath washout (MBW) is an established technique to assess functional residual capacity (FRC) and ventilation inhomogeneity in the lung. Indirect calculation of nitrogen concentration requires accurate measurement of gas concentrations. To investigate the accuracy of the CO2 concentration and molar mass (MM) values used for the indirect calculation of nitrogen concentration in a commercial MBW device [EasyOne Pro LAB (EOPL), ndd Medizintechnik AG, Switzerland] and its impact on outcomes. We used high-precision gas mixtures to evaluate CO2 and MM sensor output in vivo and in vitro. We developed updated algorithms to correct observed errors and assessed the impact on MBW outcomes and FRC measurement accuracy compared with body plethysmography. The respiratory exchange ratio (RER)-based adjustment of the measured CO2 signal used in the EOPL led to an overestimated CO2 signal (range -0.1% to 1.0%). In addition, an uncorrected dependence on humidity was identified. These combined effects resulted in an overestimation of expired nitrogen concentrations (range -0.7% to 2.6%), and consequently MBW outcomes. Corrected algorithms reduced the mean (SD) cumulative expired volume by 15.8% (9.7%), FRC by 6.6% (3.0%), and lung clearance index by 9.9% (7.6%). Differences in FRC between the EOPL and body plethysmography further increased. Inadequate signal correction causes RER- and humidity-dependent expired nitrogen concentration errors and overestimation of test outcomes. Updated algorithms reduce average signal error, however, RER values far from the population average still cause measurement errors. Despite improved signal accuracy, the updated algorithm increased the difference in FRC between the EOPL and body plethysmography.NEW & NOTEWORTHY We investigated the accuracy of the molar mass (MM) and CO2 sensors of a commercial multiple-breath washout device (ndd Medizintechnik AG, Switzerland). We identified humidity and respiratory exchange ratio-dependent errors that in most measurements resulted in an overestimation of expired nitrogen concentrations, and consequently, MBW results. Functional residual capacity and lung clearance index decreased by 6.6% and 9.9%, respectively. Despite improved signal accuracy, the difference in FRC between the EOPL and body plethysmography increased.
Related Concept Videos
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Errors occurring during blood pressure monitoring
Several factors...