Comparison of quantification of intraretinal hard exudates between optical coherence tomography en face image versus
Donghee Park1, Kyoung Lae Kim, Sung Pyo Park
1Department of Ophthalmology, Kangdong Sacred Heart Hospital, Hallym University College of Medicine, Seoul, South Korea.
Purpose:
To compare the quantification of intraretinal hard exudate (HE) using en face optical coherence tomography (OCT) and fundus photography.
Methods:
Consecutive en face images and corresponding fundus photographs from 13 eyes of 10 patients with macular edema associated with diabetic retinopathy or Coats' disease were analyzed using the machine-learning-based image analysis tool, "ilastik."
Results:
The overall measured HE area was greater with en face images than with fundus photos (en face: 0.49 ± 0.35 mm2 vs. fundus photo: 0.34 ± 0.34 mm2, P < 0.001). However, there was an excellent correlation between the two measurements (intraclass correlation coefficient [ICC] = 0.844). There was a negative correlation between HE area and central macular thickness (CMT) (r = -0.292, P = 0.001). However, HE area showed a positive correlation with CMT in the previous several months, especially in eyes treated with anti-vascular endothelial growth factor (VEGF) therapy (CMT 3 months before: r = 0.349, P = 0.001; CMT 4 months before: r = 0.287, P = 0.012).
Conclusion:
Intraretinal HE can be reliably quantified from either en face OCT images or fundus photography with the aid of an interactive machine learning-based image analysis tool. HE area changes lagged several months behind CMT changes, especially in eyes treated with anti-VEGF injections.
More Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...