Reference Ranges for Regional Cerebral Oxygen Saturation with Masimo O3 after Birth and Differences with Other
Luis Bachiller Carnicero1, Sonia Caserío Carbonero1
1Division of Neonatology, Department of Pediatrics, Rio Hortega Hospital, Valladolid, Spain.
Objective:
Cerebral oximetry using near-infrared spectroscopy (NIRS) is a noninvasive optical technology widely used in neonatology. The present study aimed to define reference ranges for cerebral tissue oxygen saturation (crSO2) with a new four-wavelength NIRS device, Masimo O3 oximeter, during immediate transition after birth and compare values with those obtained previously with NIRO 200NX®.
Study Design:
This was a prospective observational study using Masimo O3 device to measure crSO2 and regional cerebral fractional tissue oxygen extraction (cFTOE) in healthy term newborns delivered by primary cesarean section, during the 15 minutes after cord clamping. The neonates who required any medical support were excluded. The NIRS sensor was placed on the right forehead. Peripheral oxygen saturation and heart rate were continuously measured by pulse oximetry. Previous studies which established centiles for crSO2 with NIRO 200NX were used for comparison.
Results:
A total of 44 newborns were included. The median crSO2 and cFTOE (interquartile range) at 2, 5, and 7 minutes was 54% (49-54), 71% (64-86), and 79% (73-84) and 0,25 (0,18-0,33), 0,19 (0,15-0,23), and 0,16 (0,12-0,21), respectively, with no further changes afterwards. The crSO2 measurements were significantly higher with Masimo O3 compared with NIRO-200NX.
Conclusion:
The present observational study presented reference ranges for crSO2 and cFTOE measured with Masimo O3 oximeter during the immediate neonatal transition. Values obtained with O3 were higher than those obtained with other oximeters. For this reason, crSO2 is device-specific so there must be known reference values for each oximeter to define therapeutic interventions based on crSO2 and assess cerebral oxygenation in clinical studies.
Key Points:
· Masimo O3 uses four wavelengths to measure regional oxygen saturation value.. · O3 values of crSO2 and cFTOE differ with other neonatal oximeters at birth.. · Knowledge of reference range of O3 at birth is essential to guide resuscitation..
More Related Videos
11:39Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
Published on: March 14, 2013
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...