Bone resorption around the annular closure device during a postoperative follow-up of 8 years
Abdugafur Sanginov1, Aleksandr Krutko2, Olga Leonova3
1Neurosurgery Department, Novosibirsk Research Institute of Traumatology and Orthopaedics (NRITO) N.a.Ya.L.Tsivyan, Frunze Str, 17, Novosibirsk, Russia.
Objective:
Annular closure device (ACD) implantation is considered to be an effective means of preventing reherniation after microdiscectomy; however, there is an issue: the bone may resorb around the ACD. The causes of vertebral bone resorption remain unexplored; the dynamics of changes in bone resorption around the ACD have not yet been assessed or characterized.
Methods:
One hundred thirty-three patients underwent ACD implantation after microdiscectomy, and 107 of them were followed up for 8 years after surgery (Oswestry, VAS). Lumbar CT scans helped characterize the bone resorption area around the ACD.
Results:
The median of follow-up was 85 [74; 93] months (from 73 to 105 months). The prevalence of bone resorption around the ACD was up to 63.6%, and it was mainly around the polymer mesh of the ACD (70.6%). The resorbed bone volume increased with time and reached its maximum of 5.2 cm3 (12% of the vertebral body volume) once a sclerotic rim developed around the bone resorption area. No differences in VAS pain intensity or in Oswestry Disability Index were found between patients with resorption and patients without it (p > 0.05). The volume of the intervertebral disc before surgery is a predictor of bone resorption (OR = 0.79, p = 0.009): if it is less than 13.2 cm3, the risk of bone resorption increases significantly (p < 0.05).
Conclusion:
The majority of patients (up to 63.6%) with implanted ACDs have vertebral bone resorption around them. The bone resorption area around the ACD mesh increases with time to up to 12% of the vertebral body volume, with no clinical evidence, though. The formation of a sclerotic rim prevents the bone resorption area from further growth. If the volume of the intervertebral disc before surgery is less than 13.2 cm3, the risk of bone resorption increases significantly.