Stereotactic Frame-Based Electrode Insertion: The Accuracy of Increasingly Oblique Insertion Angles.
Zachary D Gilbert1, Austin M Tang, Terrance Peng
1University of Southern California, Keck School of Medicine of USC, Department of Neurological Surgery, Los Angeles, CA, United States.
Turkish Neurosurgery
|January 29, 2024
Summary
Stereotactic electrode placement accuracy is unaffected by planned drill angle. Oblique electrode insertion using the CRW frame and Ad-Tech drill guide is a safe and precise neurosurgical technique.
Area of Science:
- Neurosurgery
- Medical Devices
- Epilepsy Treatment
Background:
- Stereotactic electrode placement is crucial for treating drug-resistant epilepsy.
- The Cosman-Roberts-Wells (CRW) frame is a common tool for intracranial electrode implantation.
- Understanding the impact of planned angles on electrode accuracy is essential for surgical planning.
Purpose of the Study:
- To evaluate the relationship between planned drill approach angle and the actual angular deviation of stereotactically placed intracranial electrodes.
- To determine if the magnitude of the planned angle affects the accuracy of electrode tip placement.
Main Methods:
- Stereotactic electrode implantation in 13 patients with drug-resistant epilepsy (136 electrodes total).
- Electrode targets planned on MRI; implantation via CRW frame with Ad-Tech drill guide and electrodes.
- Post-operative CT scans used to measure actual electrode angles (axial, coronal, sagittal) and compare with planned angles via Bland-Altman plots and linear regression.
Main Results:
- Linear relationships observed between planned and actual electrode angles across axial, coronal, and sagittal planes.
- No significant angular deflection was detected for any planned angle magnitude.
- The Ad-Tech drill guide and electrodes, used with the CRW frame, demonstrated consistent performance.
Conclusions:
- The planned drill approach angle does not significantly impact the accuracy of CRW frame-based stereotactic electrode placement.
- Oblique electrode insertion is a safe and accurate neurosurgical procedure.
- Findings support the use of varied planned angles for optimal electrode targeting in epilepsy surgery.