Detecting errors in blood-gas measurement by analysis with two instruments.

L F Metzger, W B Stauffer, A V Krupinski

    Clinical Chemistry
    |April 1, 1987
    PubMed
    Summary

    Duplicate analysis effectively detects errors in blood-gas testing. This method identified more errors with older analyzers, suggesting its value for ensuring accurate blood gas analysis.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Cell-specific variant-to-gene mapping identifies conserved neural and glial regulators of sleep.

    bioRxiv : the preprint server for biology·2026
    Same author

    Estimating sleep parameters using an accelerometer without sleep diary.

    Scientific reports·2018
    Same author

    Low sensitivity of anion gap to detect clinically significant lactic acidosis in the emergency department.

    Clinical biochemistry·2017
    Same author

    Transcriptional Signatures of Sleep Duration Discordance in Monozygotic Twins.

    Sleep·2017
    Same author

    Adhesion molecule increases in sleep apnea: beneficial effect of positive airway pressure and moderation by obesity.

    International journal of obesity (2005)·2014
    Same author

    The role of obesity, different fat compartments and sleep apnea severity in circulating leptin levels: the Icelandic Sleep Apnea Cohort study.

    International journal of obesity (2005)·2012

    Area of Science:

    • Clinical Chemistry
    • Medical Diagnostics
    • Laboratory Medicine

    Background:

    • Accurate blood gas analysis is critical for patient diagnosis and management.
    • Analytical errors in blood gas measurements can lead to incorrect clinical decisions.
    • Existing error detection methods may not be sufficient to ensure reliable results.

    Purpose of the Study:

    • To evaluate the error detection capabilities of duplicate analysis for blood-gas specimens.
    • To compare the performance of different blood-gas analyzer models in error detection.
    • To assess the frequency and types of errors identified through duplicate analysis.

    Main Methods:

    • A two-stage prospective study was conducted using Corning blood-gas analyzers (Model 175 and Model 178).

    Related Experiment Videos

  • Duplicate analysis was performed on 1601 specimens in stage one and 1544 in stage two.
  • Error criteria were defined based on acceptable differences for pH, pCO2, and pO2 values.
  • Main Results:

    • Error detection rates were 6.1% (97 specimens) in stage one and 1.5% (23 specimens) in stage two.
    • More errors were detected with the older Model 175 analyzer compared to the newer Model 178.
    • Specific analytes and electrodes showed significantly higher error rates (e.g., 18% for pO2 with certain electrodes).

    Conclusions:

    • Duplicate analysis is a valuable tool for detecting errors in blood-gas specimen analysis.
    • The findings suggest that duplicate analysis could be considered a required standard for quality control.
    • Analyzer design and specific electrode performance impact error rates, highlighting the need for vigilant monitoring.