Evaluation of Sensitive Analytes to Hemolysis Interference on an Automated Chemistry Analyzer
Marfas Marakankadavu Parambu1, Valerie Bush1
1Clinical Laboratory, Bassett Medical Center, Cooperstown, NY, United States.
Background:
Hemolysis is a common reason for specimen rejection in the laboratory. Our experience suggested that hemolysis (H) flag limits are too strict for some analytes leading to unnecessary specimen rejections. This study summarizes H flags for commonly rejected analytes on the Beckman Coulter DxC 700 AU analyzer.
Methods:
We evaluated analytes with low-limit H flags and high rejection rates. These included: aspartate aminotransferase (AST), alanine aminotransferase (ALT), iron (IRN), potassium (K), direct bilirubin (DBIL), magnesium (Mg), amylase (AMY), sodium (Na), gamma-glutamyltransferase (GGT), phosphorus (PHOS), albumin (ALB), alkaline phosphatase (ALKP), and lactate dehydrogenase (LDH). Five patient plasma pools without hemolysis were made from 50 patient specimens. Neat pools were analyzed to establish baseline analyte concentrations. A hemolysate was created by diluting whole blood with distilled water. Each analyte was tested after spiking each pool with the hemolysate to specific hemoglobin concentrations corresponding to manufacturer's H flags. Percent differences were calculated between baseline pool means and each flag's pool mean. Acceptance limits were based upon the average of the 2019 CLIA and the method precision limits. Calculated percent differences greater than the acceptance limits were considered significant.
Results:
Manufacturer-defined hemolysis flags can be updated to greater than 1+ for Na, K, and AST, greater than 3+ for ALKP, and greater than 4+ for AMY and Mg. No changes were noted for the remaining analytes.
Conclusions:
The hemolysis criteria set for ALKP, AMY, AST, Mg, K, and Na were updated in the Remisol Advance middleware, which led to a 56% reduction in rejected hemolyzed specimens.
More Related Videos
Related Concept Videos
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Emission Spectroscopy: Interference
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...