Occlusion Pressures of Tactical Pneumatic Tourniquet 2"
Background:
The Tactical Pneumatic Tourniquet 2" (TPT2, 5.1cm-wide deflated) allows total average applied pressure measurement, which should be useful toward development of emergency-use limb tourniquet certification devices.
Methods:
The TPT2 hand bulb was replaced with stopcocks and syringes, allowing filling with continuous pressure measurement. Forearm and mid-thigh applications involved two sets of five Doppler-based pulse gone/return pairs. Second set pulse gones were chosen a priori for occlusion pressures (preliminary work indicated greater consistency in second sets).
Results:
All 68 forearms occluded (30 female, 38 male, median circumference 17.8cm, range 14.6-23.5cm; median second set of pulse gone tourniquet pressures 176mmHg, range 128-282mmHg). Fifty-five thighs occluded (median circumference 54.3cm, range 41.6-62.4cm; median systolic pressure 126mmHg, range 102-142mmHg; median second set of pulse gone pressures 574mmHg, range 274-1158mmHg). Thirteen thigh applications were stopped without occlusion because of concerning pressures combined with no indication of imminent occlusion and difficulties forcing more air into the TPT2 (3 female, 10 male, peak pressures from 958-1377mmHg, median 1220mmHg, p<.0001 versus occluded thighs; median circumference 63.3cm, range 55.0-72.9cm, p<.0001 versus occluded thighs; median systolic pressure 126mmHg, range 120-173mmHg, p<.019 versus occluded thighs). Thigh TPT2 impression widths on five subjects after occlusion were as follows: 3.5cm, occlusion 274mmHg; 2.8cm, occlusion 348mmHg; 2.9cm, occlusion 500mmHg; 2.8cm, occlusion 782mmHg; 2.7cm, occlusion 1114mmHg.
Conclusions:
Though probably useful to tourniquet certification, the required pressures for thigh occlusion make the TPT2 undesirable for any clinical use, emergency or otherwise.
More Related Videos
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure