Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Accuracy comparison of scan segmental sequential ranges with two intraoral scanners for maxilla and mandible
Chih-Te Liu1,2, Jen-Hao Chen1,2, Je-Kang Du1,2
1Division of Prosthodontics, Department of Dentistry, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan.
Background/Purpose:
The accuracy of a full-arch scan by using an intraoral scanner should be validated under clinical conditions. This study aimed to compare the accuracy of full-arch digital impressions in the maxilla and mandible using two intra oral scanners with three different scan segmental sequential ranges.
Materials And Methods:
A dental model with 28 teeth in their normal positions served as the reference. Sixty full-arch scans were performed using Trios 3 and Trios 4, employing scanning strategy O (manufacturer's original method), OH (segmental sequential ranges one half), and TQ (segmental sequential ranges third quarter). Trueness was evaluated by comparing digital impressions with a reference dataset using specialized software. One-way ANOVA and Tukey tests assessed differences between the groups.
Results:
For Trios 3 on the maxilla, no significant difference was found among the groups of trueness; in the mandible, strategy O exhibited a significant difference (P = 0.008) with the highest deviation. For Trios 4 on the maxilla, strategy TQ demonstrated the lowest deviation with a significant difference (P = 0.006); in the mandible, no significant difference was found among the groups of trueness.
Conclusion:
Strategy TQ exhibited the best trueness for Trios 3 and Trios 4, suggesting it may be preferred for higher accuracy. Clinicians should consider these findings when selecting scanning strategies and intraoral scanners for specific cases.