Intravoxel incoherent motion diffusion-weighted imaging of solitary pulmonary lesions: initial study with gradient-
M Yin1, Guanjie Cao2, S Lv1
1Clinical Medical College of Jining Medical University, Jining 272000, China.
Aim:
To evaluate the feasibility and image quality of intravoxel incoherent motion diffusion-weighted imaging (IVIM) using gradient- and spin-echo (GRASE) in solitary pulmonary lesions (SPLs) compared to echo planar imaging (EPI) and turbo spin-echo (TSE) at 3 T.
Materials And Methods:
Forty-two patients with SPLs underwent lung magnetic resonance imaging (MRI) using TSE-IVIM, GRASE-IVIM, and EPI-IVIM at 3 T. Signal ratio (SR), contrast ratio (CR), and image distortion ratio (DR) of three sequences were compared. The reproducibility and repeatability of the apparent diffusion coefficient (ADC) and IVIM-derived parameters were assessed using the interclass correlation coefficient (ICC) and coefficient of variation (CV). The repeatability of the ADC and IVIM-derived parameters between all sequences was evaluated using the Bland-Altman method.
Results:
EPI-IVIM had a higher SR, lower CR, and higher DR (p<0.05); however, there was no significant difference between TSE-IVIM and GRASE-IVIM (p>0.05). Compared to the D and f values of TSE-IVIM (ICC lower limit >0.90), GRASE-IVIM and EPI-IVIM showed poor reproducibility (ICC lower limit<0.90). The repeatability of the ADC and D values obtained by TSE-IVIM (CV, 1.93-2.96% and 2.44-3.18%, respectively) and GRASE-IVIM (CV, 2.56-3.12% and 3.21-3.51%, respectively) were superior to those of EPI-IVIM (CV, 10.03-10.2% and 11.30-11.57%). The repeatability of D∗ and f values for all sequences was poor. Bland-Altman analysis showed wide limits of agreement between the ADC and IVIM-derived parameters for all sequences.
Conclusion:
GRASE-IVIM reduced the DR, improved the stability of the ADC and D values on repeated scans, and had the shortest scanning time.