Related Experiment Video
Updated: May 1, 2026

Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Are offline ART decisions for NSCLC impacted by the type of dose calculation algorithm?
Dylan Callens1,2, Karel Aerts1, Patrick Berkovic1,2
1Laboratory of Experimental Radiotherapy, KU Leuven, Leuven, Belgium.
Introduction:
Decisions for plan-adaptations may be impacted by a transitioning from one dose-calculation algorithm to another. This study examines the impact on dosimetric-triggered offline adaptation in LA-NSCLC in the context of a transition from superposition/convolution dose calculation algorithm (Type-B) to linear Boltzmann equation solver dose calculation algorithms (Type-C).
Materials & Methods:
Two dosimetric-triggered offline adaptive treatment workflows are compared in a retrospective planning study on 30 LA-NSCLC patients. One workflow uses a Type-B dose calculation algorithm and the other uses Type-C. Treatment plans were re-calculated on the anatomy of a mid-treatment synthetic-CT utilizing the same algorithm utilized for pre-treatment planning. Assessment for plan-adaptation was evaluated through a decision model based on target coverage and OAR constraint violation. The impact of algorithm during treatment planning was controlled for by recalculating the Type-B plan with Type-C.
Results:
In the Type-B approach, 13 patients required adaptation due to OAR-constraint violations, while 15 patients required adaptation in the Type-C approach. For 8 out of 30 cases, the decision to adapt was opposite in both approaches. None of the patients in our dataset encountered CTV-target underdosage that necessitated plan-adaptation. Upon recalculating the Type-B approach with the Type-C algorithm, it was shown that 10 of the original Type-B plans revealed clinically relevant dose reductions (≥3%) on the CTV in their original plans. This re-calculation identified 21 plans in total that required ART.
Discussion:
In our study, nearly one-third of the cases would have a different decision for plan-adaption when utilizing Type-C instead of Type-B. There was no substantial increase in the total number of plan-adaptations for LA-NSCLC. However, Type-C is more sensitive to altered anatomy during treatment compared to Type-B. Recalculating Type-B plans with the Type-C algorithm revealed an increase from 13 to 21 cases triggering ART.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant