Realtime monitoring of thrombus formation in vivo using a self-reporting vascular access graft

Daniel Hoare1, David Kingsmore2, Michael Holsgrove3

  • 1School of Cardiovascular & Metabolic Health, University of Glasgow, Glasgow, UK.

Communications Medicine
|February 5, 2024
PubMed

Insights

A new system uses electrical impedance spectroscopy sensors and radiotelemetry to monitor synthetic arteriovenous grafts. This technology enables real-time, wireless detection of blockages, potentially preventing graft failure in dialysis patients.

Area of Science:

  • Biomedical Engineering
  • Cardiovascular Research
  • Medical Devices

Background:

  • Chronic kidney disease (CKD) impacts 10% of the global population, with significant cardiovascular risks.
  • Synthetic arteriovenous grafts (AVGs) are crucial for dialysis access but are prone to neointimal hyperplasia, leading to thrombosis and failure.

Purpose of the Study:

  • To develop and evaluate a novel system for real-time, in vivo monitoring of synthetic arteriovenous grafts.
  • To enable early detection of complications like venous stenosis and thrombosis in AVGs.

Main Methods:

  • Integration of electrical impedance spectroscopy (EIS) sensors into the AVG wall.
  • Utilization of an implantable radiotelemetry system for wireless data transmission.
  • In vitro testing with cell cultures and blood, followed by in vivo implantation in swine.

Main Results:

  • The system successfully detected and wirelessly reported the accumulation of biological material (cells, blood).
  • Distinguished between normal and pathological conditions, including atheroma.
  • Remotely monitored graft blockage formation in swine.

Conclusions:

  • Combining EIS and radiotelemetry offers a promising approach for continuous graft surveillance.
  • Enables early detection of venous stenosis and blood clot formation in real-time.
  • Potential applications extend to other cardiovascular implants and diseases.
Abstract