Woo Jung Seo1, Eun Young Kim2, Ga Jin Seo2
1Division of Pulmonary and Critical Care Medicine, Department of Internal Medicine, Ilsan Paik Hospital, Inje University College of Medicine, Goyang, Republic of Korea.
Background:
Patients with respiratory disease often need oxygen supplements through a High-flow nasal cannula (HFNC), both hypoxia and hyperoxia can be harmful. Proper oxygen therapy requires careful monitoring of oxygen levels and adjustments to oxygen levels. A new automated system called Targeted SpO2 Feedback (TSF) improved the oxygen delivery system compared to manual adjustments. [Correction added on 25 October 2024, after first online publication: Background subsection in Abstract has been added on this version.] AIM: To test whether targeted SpO2 feedback (TSF), an automatic control system for fraction of inspired oxygen (FiO2), achieves more time in the optimal SpO2 range and/or reduces the frequency of manual adjustments to administered FiO2 compared with conventional manual titration in patients with hypoxia on high-flow nasal cannula (HFNC) therapy.
Study Design:
Twenty-two patients were recruited from two hospitals. For each, two sessions of manual mode and two sessions of TSF were applied in a random order, each session lasting 2 h. The target SpO2 on TSF was 95%. Oxygen monitoring levels were classified into four SpO2 ranges: hypoxia (≤ 89%), borderline (90%-93%), optimal (94%-96%) and hyperoxia (≥ 97%). The two modes were compared based on the proportion of time spent in each SpO2 range and the number of manual FiO2 adjustments.
Results:
The proportion of time in the optimal SpO2 range was 20.5% under manual titration mode and 65.4% under TSF (p < .01). The proportions of time in the hypoxia range were 1.1% and 0.4%, respectively (p = .31), in the borderline range 4.7% and 3.5%, respectively (p = .54), and in the hyperoxia range 73.7% and 30.7%, respectively (p < .01). There were statistical differences only in the optimal and hyperoxia SpO2 ranges. During the 8 h, the frequency of manual FiO2 adjustment was 0.7 times for the manual mode and 0.2 times for TSF, showing no statistically significant difference (p = 0.076).
Conclusion:
Compared with manual titration, TSF achieved greater time of the optimal SpO2 and less time of hyperoxia during HFNC. The frequency of manual adjustments on TSF tended to be less than on manual titration mode.
Relevance To Clinical Practice:
Automatic closed-loop algorithm FiO2 monitoring systems can achieve better oxygen treatments than conventional monitoring and may reduce nurse workloads. In the era of pandemic respiratory diseases, this system can also facilitate contactless SpO2 monitoring during HFNC therapy.
More Related Videos
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...