Related Experiment Video
Updated: May 8, 2026

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
A zero-exchange approach for left atrial access in pulmonary vein isolation with pulsed field ablation
Alexandru Gabriel Bejinariu1, Maximilian Spieker1, Hisaki Makimoto2
1Division of Cardiology, Pulmonology and Vascular Medicine, Medical Faculty, Heinrich-Heine University, Düsseldorf, Germany.
Introduction:
Pulsed field ablation (PFA) has emerged as an innovative technique for pulmonary vein isolation (PVI). Typically, a transeptal puncture (TSP) with a standard sheath precedes a switch to the larger diameter sheath in the left atrium. This study aimed to describe the safety and feasibility of direct TSP using the large diameter Faradrive sheath before performing PVI with PFA.
Methods:
We prospectively enrolled 166 consecutive patients with paroxysmal or persistent atrial fibrillation (AF) undergoing PVI with PFA at our institution. TSP was performed in all cases with transesophageal echocardiography guidance, using the Faradrive sheath and a 98 cm matched Brockenbrough needle. The primary endpoint was the occurrence of pericardial tamponade during or within the first 48 h after the procedure. The secondary endpoint was the occurrence of any major complication.
Results:
All 166 patients were included into the final analysis (44% female): 64% of patients had paroxysmal AF and 36% persistent AF (68 ± 11 years old, median CHA2DS2Vasc Score 3, median left atrial volume index 31). The median duration of the procedure was 60 min, median time to TSP was 15 min, and the median fluoroscopy dose was 595 cGy × cm2. The primary endpoint occurred in one patient: a non-TSP related pericardial tamponade, which was managed with pericardial puncture.
Conclusion:
Direct TSP with skipping sheath exchange using the large diameter Faradrive sheath for PVI with PFA was safe, feasible, and reduced costs in all patients. Large scale studies and registries are needed to verify this workflow.
Related Concept Videos
Chambers of the Heart
Deoxygenated blood from the body is received in the right...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization