The effect of ablation settings on lesion characteristics with DiamondTemp ablation system: An ex vivo experiment
Takehiro Nomura1, Manabu Maeda1, Daiki Kumazawa1
1Heart Rhythm Center, Department of Cardiovascular Medicine Sendai Kosei Hospital Sendai-shi Miyagi Japan.
Introduction:
Creating large lesion in ablations using the DiamondTemp (DTA) ablation system may reduce the frequency of arrhythmia recurrence and allow the treatment of ventricular arrhythmias. Therefore, this study aimed to investigate whether power, application time, contact force (CF), and contact angle affect lesion formation in the ventricles.
Methods:
Ablations were delivered to porcine myocardial preps to evaluate the lesion characteristics. Ablations were conducted with a maximum power of 50 W, target temperature of 58°C, CF of 10, 20, or 30 g, and contact angle between the catheter tip and tissue. The ablation durations were 15, 30, 60 s, 15 s × 2, or 30 s × 2.
Results:
Steam pops occurred only in cases with perpendicular contact. The lesion depth was larger in all settings in the perpendicular orientation than in the parallel orientation. The temperatures were lower in all settings in the perpendicular orientation than in the parallel orientation. The lesions became larger as CF increased with perpendicular contact and duration of ≥30 s. The longer application time resulted in larger surface area, depth, and volume of the lesion. Lesion depth was greater with single application of 30 and 60 s than with 15 s × 2 and 30 s × 2, respectively.
Conclusion:
It is important to perform a single prolonged application as much as possible to create deeper lesions. Parallel contact with the tissue should be maintained to take advantage of the temperature sensor's capabilities to avoid pop phenomenon.