Related Experiment Video
Updated: Jul 4, 2025

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Zero to minimal fluoroscopy for cardiac electronic device implantation: A systematic review and meta-analysis
Kevin Luke1, Clonia Milla1, Joshua Kurnia Tandi2
1Faculty of Medicine Universitas Airlangga Surabaya Indonesia.
Background:
Fluoroscopy is conventionally performed for cardiac implantable electronic device (CIED) therapy and carries radiation drawback for both patients and medical workers. Recently, zero to minimal fluoroscopy (ZMF) approach is introduced to reduce radiation exposure of fluoroscopy. This study compares the feasibility and safety of ZMF approach to fluoroscopy for CIEDs therapy in adults.
Method:
A systematic literature search was conducted on PubMed, ScienceDirect, and Web of Science in March 2023. All observational or experimental studies comparing ZMF approach to fluoroscopy for adult CIEDs therapy were included. Reviews, case report/series, animal studies, and non-English articles were excluded. The success rate, procedural time, fluoroscopy time, radiation dose, and complications rate were compared for each approach.
Results:
Seven articles for permanent and three articles for temporary CIEDs were included for analysis. The success rate of ZMF for permanent CIEDs was similar to fluoroscopy method (OR: 0.77, 95% CI: 0.33-4.15). The procedural time of ZMF was similar to fluoroscopy for both permanent and temporary CIEDs (standardized mean difference [SMD]: 0.10, 95% CI: -0.35 to 0.55 and SMD: -0.71, 95% CI: -1.87-0.44, respectively). However, ZMF approach markedly reduced the fluoroscopy time and radiation exposure for permanent CIEDs (SMD: -1.80, 95% CI: -2.49 to -1.12 and SMD: -1.26, 95% CI: -2.24 to -0.29). The complication rate was similar for permanent CIEDs (OR: 1.08, 95% CI: 0.41-2.84), yet lowered for temporary CIEDs (OR: 0.34, 95% CI: 0.20-0.59).
Conclusion:
ZMF had similar success rate, procedural time, and sum complication rate for permanent CIEDs implantation with a significant reduction of fluoroscopy time and radiation exposure.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...