UHPLC-MS/MS for plasma lamotrigine analysis and comparison with a homogenous enzyme immunoassay
Xiaoxu Shi1, Dongjie Zhang2,3, Zhigang Zhao2,3
1Children's Hospital, Capital Institute Of Paediatrics, 2 Yabao Road, Chaoyang District, Beijing, 100020, China.
Abstract:
Aims: To develop and validate a UHPLC-MS/MS method for lamotrigine (LTG) analysis in human plasma and evaluate its agreement with a homogenous enzyme immunoassay (HEIA). Materials & methods: The UHPLC-MS/MS method was developed and validated according to the USFDA/EMA guidelines. A Bland-Altman plot was used to evaluate the agreement between UHPLC-MS/MS and HEIA. Results: Samples were pretreated with one-step protein precipitation and separated in 2.6 min. The intra- and inter-day bias and imprecisions were -15.8 to 15.0% and less than 11.17%, respectively. The recovery and matrix factor were 98.30 to 111.97%. The mean overestimation of UHPLC-MS/MS compared with HEIA was 21.57%. Conclusion: A rapid, sensitive and robust UHPLC-MS/MS method for plasma LTG analysis was developed and validated and was a 21.57% overestimation compared with HEIA.