The Edwards Acumen IQ system using peripheral arterial catheter-based waveforms to estimate cardiac output is not
Objective:
To assess the agreement between cardiac output (CO) estimated via evaluation of the arterial pressure waveform by a novel monitoring system (Edwards Acumen IQ sensor and HemoSphere Advanced Monitor Platform [HS-IQ]; Edwards LifeSciences) and measured by thermodilution (TD) in anesthetized, normovolemic, and hypovolemic dogs. To assess the agreement between the HS-IQ CO measurements in the radial artery and dorsal metatarsal artery.
Animals:
8 purpose-bred Beagles.
Methods:
Dogs were placed under general anesthesia. CO was measured via TD and via the HS-IQ at radial and dorsal metatarsal arterial catheters. CO measurements were obtained at 4 time points including normovolemic and multiple hypovolemic states. Paired measurements of CO were evaluated via the method of Bland and Altman with acceptable limits of agreement (LOA) defined as < 30%.
Results:
A total of 24 (dorsal metatarsal) and 21 (radial) paired measurements were collected in 8 dogs. The overall bias (CI) for comparison of TD to radial arterial HS-IQ CO measurements was -0.09 L/min. LOA and proportional LOA were -2.66 to 2.49 L/min and -140.72% to 104.94%. The overall bias (CI) for comparison of TD to dorsal metatarsal arterial HS-IQ CO measurements was -0.26 L/min. LOA and proportional LOA were -2.76 to 2.24 L/min and -135.96% to 93.25%. The overall proportional error for radial arterial was -17.9% and for dorsal metatarsal was -21.4%.
Clinical Relevance:
CO measurements with the HS-IQ were easy to obtain but did not produce results within a clinically acceptable range for either measurement site, with a very wide LOA. The CO estimations from the HS-IQ are not appropriate for clinical use at this time.
More Related Videos
07:51Standardized Hemorrhagic Shock Induction Guided by Cerebral Oximetry and Extended Hemodynamic Monitoring in Pigs
Published on: May 21, 2019
06:10Author Spotlight: Assessment of Cardiac Output Calculation by Thermodilution in Pigs for Effective Perfusion Flow During EVLP
Published on: June 28, 2024
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation