Ionizing radiation exposure in complex percutaneous coronary intervention: Defining local diagnostic reference levels
Hugo Costa1, Miguel Espírito-Santo1, João Bispo1
1Cardiology Department, Centro Hospitalar Universitário do Algarve, Faro, Portugal.
Introduction And Objectives:
Concerns regarding the consequences of ionizing radiation (IR) have been increasing in the field of interventional cardiology (IC). There is little information on reported national and local radiation diagnostic reference levels (DRLs) in catheterization laboratories in Portugal. This study was designed to assess the IR dose exposure during complex percutaneous coronary intervention (PCI), and to set the respective DRLs and future achievable doses (ADs).
Methods:
This was a retrospective cohort study which took place between 2019 and 2020, including patients who underwent complex PCI. Complex PCI was defined as all procedures that encompass treatment of chronic total occlusions (CTO) or left main coronary artery. DRLs were defined as the 75th percentile of the distribution of the median values of air kerma area product (PKA) and cumulative air kerma (Ka,r). ADs were set at the 50th percentile of the study dose distribution. Multivariate analysis was performed using linear regression to identify predictors significantly associated with radiation dose (Ka,r).
Results:
A total of 242 patients were included in the analysis. Most patients underwent a CTO procedure (146, 60.3%). Patients were aged 67.9±11.2 years and mostly male (81.4%). DRLs were set in Ka,r (3012 mGy) and PKA (162 Gy cm2) for complex PCI. ADs were also set in Ka,r (1917 mGy) and PKA (101 Gy cm2). Independent predictors of Ka,r with a positive correlation were PKA (0.893, p<0.001), fluoroscopy time (0.520, p<0.001) and PCI time (0.521, p<0.001).
Conclusions:
This study reports the results of IR in complex PCI. DRLs were set for IR dose exposure measured in Ka,r (3012 mGy) and PKA (162 Gy cm2). ADs, values to be achieved in future assessment, were set to Ka,r (1917 mGy) and PKA (101 Gy cm2).
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT