Dose Calculation Accuracy of Beam Models in RadCalc for a 1.5 T MR-Linac
Jiwon Sung1, Yeonho Choi1, Jun Won Kim1
1Department of Radiation Oncology, Gangnam Severance Hospital, Yonsei University College of Medicine, Seoul 06273, Republic of Korea.
RadCalc software version 7.1.4 accurately verifies patient-specific quality assurance (PSQA) doses for online adaptive radiotherapy using a 1.5 T magnetic resonance linear accelerator (MR-linac). This independent dose calculation tool is reliable for adaptive plans.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Quality Assurance
Background:
- Online adaptive radiotherapy requires precise dose verification.
- Magnetic resonance linear accelerators (MR-linacs) present unique challenges for dose calculation due to magnetic field interactions.
- Independent software is crucial for patient-specific quality assurance (PSQA).
Purpose of the Study:
- To evaluate RadCalc version 7.1.4 for PSQA in 1.5 T MR-linac online adaptive planning.
- To assess the accuracy of RadCalc's beam modeling incorporating MR field characteristics.
- To validate RadCalc as a reliable tool for adaptive radiotherapy dose verification.
Main Methods:
- Established six RadCalc beam models using various percentage depth dose (PDD) data sources.
- Assessed dose-calculation accuracy using simple tests on a virtual water phantom.
- Conducted clinical tests on 575 adaptive plans (prostate, liver, breast) comparing RadCalc to PSQA measurements.
Main Results:
- Dose calculation differences ranged from -0.5% to -20.6% in virtual phantom tests.
- Clinical tests on adaptive plans showed percentage differences of -0.51% (prostate), 1.12% (liver), and 4.10% (breast).
- RadCalc calculations demonstrated similar trends to PSQA-based measurements.
Conclusions:
- RadCalc version 7.1.4 successfully models 1.5 T magnetic field characteristics for adaptive planning.
- The software's dose calculation accuracy at 1.5 T has been verified.
- RadCalc is a reliable and effective tool for PSQA in MR-linac adaptive radiotherapy.