Can Digital Enhancement Restore the Image Quality of Phosphor Plate-Based Radiographs Partially Damaged by Ambient
Matheus Sampaio-Oliveira1, Luiz Eduardo Marinho-Vieira2, Matheus Barros-Costa2
1Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas (UNICAMP), Av. Limeira, 901, Piracicaba-SP, 13414-903, Brazil. m234718@dac.unicamp.br.
Abstract:
To assess the effect of digital enhancement on the image quality of radiographs obtained with photostimulable phosphor (PSP) plates partially damaged by ambient light. Radiographs of an aluminum step wedge were obtained using the VistaScan and Express systems. Half of the PSP plates was exposed to ambient light for 0, 10, 30, 60, or 90 s before being scanned. The resulting radiographs were exported with and without digital enhancement. Metrics for brightness, contrast, and contrast-to-noise ratio (CNR) were derived, and the ratio of each metric between the exposed-to-light and non-exposed-to-light halves of the radiographs was calculated. The resulting ratios of the radiographs with digital enhancement were subtracted from those without digital enhancement and compared among each other. For the VistaScan system, digital enhancement partially restored brightness, contrast, and CNR. For the Express system, digital enhancement only restored CNR and not the impact of ambient light on brightness and contrast. Specifically, digital enhancement restored 23.48% of brightness for the VistaScan, while percentages below 1% were observed for the Express. Digital enhancement restored 53.25% of image contrast for the VistaScan and 5.79% for the Express; 40.71% of CNR was restored for the VistaScan, and 35% for the Express. Digital enhancement can partially restore the damage caused by ambient light on the brightness and contrast of PSP-based radiographs obtained with the VistaScan, as well as on CNR for the VistaScan and Express systems. The exposure of PSP plates to light can lead to unnecessary retakes and increased patient exposure to X-rays.
More Related Videos
09:34Semi-Automated Planimetric Quantification of Dental Plaque Using an Intraoral Fluorescence Camera
Published on: January 27, 2023
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
Related Concept Videos
X-ray Imaging
Photoluminescence: Applications
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Photoreceptors and Visual Pathways
Photoluminescence: Fluorescence and Phosphorescence
A pair of electrons in a...