Related Experiment Video
Updated: Jul 3, 2025

Dynamic Quantitative Sensory Testing to Characterize Central Pain Processing
Published on: February 16, 2017
Optimizing Temporal Summation of Heat Pain Using a Constant Contact Heat Stimulator
Parker A Kell1, Claudia N Vore1, Burkhart J Hahn1,2
1Department of Psychology, The University of Tulsa, Tulsa, OK, USA.
Purpose:
Temporal summation (TS) of pain occurs when pain increases over repeated presentations of identical noxious stimuli. TS paradigms can model central sensitization, a state of hyperexcitability in nociceptive pathways that promotes chronic pain onset and maintenance. Many experimenters use painful heat stimuli to measure TS (TS-heat); yet, TS-heat research faces unresolved challenges, including difficulty evoking summation in up to 30-50% of participants. Moreover, substantial variability exists between laboratories regarding the methods for evoking and calculating TS-heat.
Patients And Methods:
To address these limitations, this study sought to identify optimal parameters for evoking TS-heat in healthy participants with a commercially available constant contact heat stimulator, the Medoc TSA-II. Working within constraints of the TSA-II, stimulus trains with varying parameters (eg, stimulus frequency, baseline temp, peak temp, peak duration, testing site) were tested in a sample of 32 healthy, chronic pain-free participants to determine which combination best evoked TS-heat. To determine whether TS scoring method altered results, TS-heat was scored using three common methods.
Results:
Across all methods, only two trains successfully evoked group-level TS-heat. These trains shared the following parameters: site (palmar hand), baseline and peak temperatures (44°C and 50°C, respectively), and peak duration (0.5 s). Both produced summation that peaked at moderate pain (~50 out of 100 rating).
Conclusion:
Future TS-heat investigations using constant contact thermodes and fixed protocols may benefit from adopting stimulus parameters that include testing on the palmar hand, using 44°C baseline and 50°C peak temperatures, at ≥0.33 Hz stimulus frequency, and peak pulse durations of at least 0.5 seconds.