Modified Theoretical Model Predicts Radial Support Capacity of Polymer Braided Stents
Xue Hu1, Qingwei Liu1, Li Chen2
1School of Mechanical Engineering, Jiangsu Key Laboratory for Design and Manufacture of Micro-Nano Biomedical Instruments, Southeast University, Nanjing 211189, Jiangsu, China.
Background And Objective:
Self-expanding polymer braided stents are expected to replace metallic stents in the treatment of Peripheral Arterial Disease, which seriously endangers human health. To restore the patency of blocked peripheral arteries with different properties and functions, the radial supporting capacity of the stent should be considered corresponding to the vessel. A theoretical model can be established as an effective method to study the radial supporting capacity of the stent which can shorten the stent design cycle and realize the customization of the stent according to lesion site. However, the classical model developed by Jedwab and Clerc of radial force is only limited to metallic braided stents, and the predictions for polymer braided stents are deviated.
Methods:
In this paper, based on the limitation of the J&C model for polymer braided stents, a modified radial force model for polymer braided stents was proposed, which considered the friction between monofilaments and the torsion of the monofilaments. And the modified model was verified by radial force tests of polymer braided stents with different structures and monofilaments.
Results:
Compared with the J&C model, the proposed modified model has better predictability for the radial force of polymer braided stents that prepared with different braided structure and polymer monofilaments. The root mean squared error of modified model is 0.041±0.026, while that of the J&C model is 0.246±0.111.
Conclusions:
For polymer braided stents, the friction between the polymer monofilaments and the torsion of the monofilaments during the radial compression cannot be ignored. The radial force prediction accuracy of the modified model considering these factors was significantly improved. This work provides a research basis on the theoretical model of polymer braided stents, and improves the feasibility of rapid personalized customization of polymer braided stents.
More Related Videos
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
10:47Tri-layered Electrospinning to Mimic Native Arterial Architecture using Polycaprolactone, Elastin, and Collagen: A Preliminary Study
Published on: January 4, 2011
Related Concept Videos
Circular Shafts - Elastoplastic Materials
As torque on the...
Members Made of Elastoplastic Material
As the bending moment...
Plastic Deformation in Circular Shafts
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Stress Concentrations in Circular Shafts
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity