An intuitive approach for spike removal in Raman spectra based on peaks' prominence and width
1Instituto de Catálisis y Petroleoquímica (ICP), CSIC, Marie Curie, 2, Madrid, 28049, Spain.
Background:
Raman spectroscopists are familiar with the challenge of dealing with spikes caused by cosmic rays. These artifacts may lead to errors in subsequent data processing steps, such as for example calibration, normalization or spectral search. Spike removal is therefore a fundamental step in Raman spectral data pre-treatment, but access to publicly accessible code for spike removal tools is limited, and their performance for spectra correction often unknown. Therefore, there is a need for development and testing open-source and easy-to-implement algorithms that improve the Raman data processing workflow.
Results:
In this work, we present and validate two approaches for spike detection and correction in Raman spectral data from graphene: i) An algorithm based on the peaks' widths and prominences and ii) an algorithm based on the ratio of these two peak features. The first algorithm provides an efficient and reliable approach for spike detection in real and synthetic Raman spectra by imposing thresholds on the peaks' width and prominence. The second approach leverages the prominence/width ratio for outlier detection. It relies on the calculation of a limit of detection based on either one or several spectra, enabling the automatic identification of cosmic ray and low-intensity noise-originated spikes alike. Both algorithms detect low-intensity spikes down to at least ≈10% of the highest Raman peak of spectra with different noise levels. To address their limitations and prove their versatility, the algorithms were further tested in Raman spectra from calcite and polystyrene.
Significance:
Our intuitive, open-source algorithms have been validated and allow automatic correction for a given set of samples. They do not require any pre-processing steps such as calibration or baseline subtraction, and their implementation with Python libraries is computationally efficient, allowing for immediate utilization within existing open-source packages for Raman spectra processing.
More Related Videos
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
¹H NMR Signal Integration: Overview
NMR Spectrometers: Resolution and Error Correction