Phantom study and clinical application of total-body 18F-FDG PET/CT imaging: How to use small voxel imaging better?
Chi Qi1,2,3,4,5, Xiuli Sui1,2,3,4, Haojun Yu1,2,3,4
1Department of Nuclear Medicine, Zhongshan Hospital, Fudan University, No. 180 in Fenglin Road, Shanghai, 200032, People's Republic of China.
Background:
Conventional PET/CT imaging reconstruction is typically performed using voxel size of 3.0-4.0 mm in three axes. It is hypothesized that a smaller voxel sizes could improve the accuracy of small lesion detection. This study aims to explore the advantages and conditions of small voxel imaging on clinical application.
Methods:
Both NEMA IQ phantom and 30 patients with an injected dose of 3.7 MBq/kg were scanned using a total-body PET/CT (uEXPLORER). Images were reconstructed using matrices of 192 × 192, 512 × 512, and 1024 × 1024 with scanning duration of 3 min, 5 min, 8 min, and 10 min, respectively.
Results:
In the phantom study, the contrast recovery coefficient reached the maximum in matrix group of 512 × 512, and background variability increased as voxel size decreased. In the clinical study, SUVmax, SD, and TLR increased, while SNR decreased as the voxel size decreased. When the scanning duration increased, SNR increased, while SUVmax, SD, and TLR decreased. The SUVmean was more reluctant to the changes in imaging matrix and scanning duration. The mean subjective scores for all 512 × 512 groups and 1024 × 1024 groups (scanning duration ≥ 8 min) were over three points. One false-positive lesion was found in groups of 512 × 512 with scanning duration of 3 min, 1024 × 1024 with 3 min and 5 min, respectively. Meanwhile, the false-negative lesions found in group of 192 × 192 with duration of 3 min and 5 min, 512 × 512 with 3 min and 1024 × 1024 with 3 min and 5 min were 5, 4, 1, 4, and 1, respectively. The reconstruction time and storage space occupation were significantly increased as the imaging matrix increased.
Conclusions:
PET/CT imaging with smaller voxel can improve SUVmax and TLR of lesions, which is advantageous for the diagnosis of small or hypometabolic lesions if with sufficient counts. With an 18F-FDG injection dose of 3.7 MBq/kg, uEXPLORER PET/CT imaging using matrix of 512 × 512 with 5 min or 1024 × 1024 with 8 min can meet the image requirements for clinical use.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
![Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F3777.jpg&w=3840&q=50)