Clinical Super-Resolution Computed Tomography of Bone Microstructure: Application in Musculoskeletal and Dental
Santeri J O Rytky1, Aleksei Tiulpin2,3, Mikko A J Finnilä2,4
1Research Unit of Health Sciences and Technology, University of Oulu, POB 5000, 90014, Oulu, Finland. santeri.rytky@oulu.fi.
Purpose:
Clinical cone-beam computed tomography (CBCT) devices are limited to imaging features of half a millimeter in size and cannot quantify the tissue microstructure. We demonstrate a robust deep-learning method for enhancing clinical CT images, only requiring a limited set of easy-to-acquire training data.
Methods:
Knee tissue from five cadavers and six total knee replacement patients, and 14 teeth from eight patients were scanned using laboratory CT as training data for the developed super-resolution (SR) technique. The method was benchmarked against ex vivo test set, 52 osteochondral samples are imaged with clinical and laboratory CT. A quality assurance phantom was imaged with clinical CT to quantify the technical image quality. To visually assess the clinical image quality, musculoskeletal and maxillofacial CBCT studies were enhanced with SR and contrasted to interpolated images. A dental radiologist and surgeon reviewed the maxillofacial images.
Results:
The SR models predicted the bone morphological parameters on the ex vivo test set more accurately than conventional image processing. The phantom analysis confirmed higher spatial resolution on the SR images than interpolation, but image grayscales were modified. Musculoskeletal and maxillofacial CBCT images showed more details on SR than interpolation; however, artifacts were observed near the crown of the teeth. The readers assessed mediocre overall scores for both SR and interpolation. The source code and pretrained networks are publicly available.
Conclusion:
Model training with laboratory modalities could push the resolution limit beyond state-of-the-art clinical musculoskeletal and dental CBCT. A larger maxillofacial training dataset is recommended for dental applications.
More Related Videos
09:36Outer-Boundary Assisted Segmentation and Quantification of Trabecular Bones by an Imagej Plugin
Published on: March 14, 2018
06:59Author Spotlight: An Economic and Efficient Method for Quantitative Evaluation of Bone Microarchitecture in a Murine Osteoporosis Model
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...