Stability and Analytical Characterization of Voriconazole as Measured by Immunoassay
Veni Bharti1, Benjamin Jung1,2, Sarah Lam3
1Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada.
Background:
Voriconazole is a broad-spectrum triazole antifungal agent recommended for invasive fungal diseases, including invasive aspergillosis. Therapeutic drug monitoring via voriconazole target trough concentration is important to ensure efficacy while preventing toxicity. Our aim was to determine the stability of voriconazole as adapted and measured by an immunoassay.
Methods:
Plasma from patient samples (n = 45) evaluated by a liquid chromatography with tandem mass spectrometry (LC-MS/MS) method was compared against an ARK immunoassay method, adapted and optimized on the Abbott Alinity c analyzer. Stability of voriconazole and analytical performance of ARK immunoassay was assessed, including functional sensitivity, limit of blank (LoB), limit of detection (LoD), and limit of quantification (LoQ), linearity, and precision.
Results:
ARK voriconazole immunoassay was highly correlated (Pearson R = 0.988) to the LC-MS/MS method, with an average bias of 0.09 mg/L (2%). CV at LoQ of 0.5 mg/L was 3.7% while the functional sensitivity was established at 0.05 mg/L. Overall imprecision with liquid quality control material obtained from ARK was 5.0%, 6.3%, and 5.9% at 1 mg/L, 5 mg/L, and 10 mg/L, respectively. Limit of blank and LoD were 0.02 mg/L and 0.05 mg/L, respectively. Voriconazole in lithium heparin plasma separator tube declines over time, with a decrease that is more evident near or above toxic concentrations.
Conclusion:
Voriconazole collected in gel separation tubes declines over time, possibly due to absorptive properties. Voriconazole measurements by immunoassay and LC-MS/MS demonstrated acceptable comparability with sufficient level of sensitivity and precision.
More Related Videos
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the lowest drug...
Measurement of Bioavailability: Pharmacokinetic Methods
Measurement of Bioavailability: Pharmacodynamic Methods
Pharmaceutical Alternatives: Stability-Related Therapeutic Nonequivalence
Therapeutic Drug Monitoring: Drug Analysis Methods