Objective:
Radiation exposure during complex endovascular aortic repair may be associated with tangible adverse effects in patients and operators. This study aimed to identify the steps of highest radiation exposure during fenestrated endovascular aortic repair (FEVAR) and to investigate potential intraoperative factors affecting radiation exposure.
Methods:
Prospective data of 31 consecutive patients managed exclusively with four-fenestration endografts between March 1, 2020, and July 1, 2022 were retrospectively analyzed. Leveraging the conformity of the applied technique, every FEVAR operation was considered a combination of six overall stages composed of 28 standardized steps. Intraoperative parameters, including air kerma, dose area product, fluoroscopy time, and number of digital subtraction angiographies (DSAs) and average angulations were collected and analyzed for each step.
Results:
The mean procedure duration and fluoroscopy time was 140 minutes (standard deviation [SD], 32 minutes), and 40 minutes (SD, 9.1 minutes), respectively. The mean air kerma was 814 mGy (SD, 498 mGy), and the mean dose area product was 66.8 Gy cm2 (SD, 33 Gy cm2). The percentage of air kerma of the entire procedure was distributed throughout the following procedure stages: preparation (13.9%), main body (9.6%), target vessel cannulation (27.8%), stent deployment (29.1%), distal aortoiliac grafting (14.3%), and completion (5.3%). DSAs represented 23.0% of the total air kerma. Target vessel cannulation and stent deployment presented the highest mean lateral angulation (67 and 63 degrees, respectively). Using linear regression, each minute of continuous fluoroscopy added 18.9 mGy of air kerma (95% confidence interval, 17.6-20.2 mGy), and each DSA series added 21.1 mGy of air kerma (95% confidence interval, 17.9-24.3 mGy). Body mass index and lateral angulation were significantly associated with increased air kerma (P < .001).
Conclusions:
Cannulation of target vessels and bridging stent deployment are the steps requiring the highest radiation exposure during FEVAR cases. Optimized operator protection during these steps is mandatory.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
06:26Novel and Innovative Hybrid Technique for Type A Aortic Dissection
Published on: March 28, 2025
Related Concept Videos
Aneurysm III: Interprofessional Care
Aneurysm IV: Nursing Management