In vivo validation of an in situ calibration bead as a reference for backscatter coefficient calculation
Yuning Zhao1,2, Gregory J Czarnota3, Trevor H Park4
1Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, IL 61801.
Objectives:
The study aims to assess the capability of Quantitative Ultrasound (QUS) based on the backscatter coefficient (BSC) for classifying disease states, such as breast cancer response to neoadjuvant chemotherapy and quantifying fatty liver disease. We evaluate the effectiveness of an in situ titanium (Ti) bead as a reference target in calibrating the system and mitigating attenuation and transmission loss effects on BSC estimation.
Methods:
Traditional BSC estimation methods require external references for calibration, which do not account for ultrasound attenuation or transmission losses through tissues. To address this issue, we use an in situ titanium (Ti) bead as a reference target, because it can be used to calibrate the system and mitigate the attenuation and transmission loss effects on estimation of the BSC. The capabilities of the in situ calibration approach were assessed by quantifying consistency of BSC estimates from rabbit mammary tumors (). Specifically, mammary tumors were grown in rabbits and when a tumor reached 1 cm or greater in size, a 2-mm Ti bead was implanted into the tumor as a radiological marker and a calibration source for ultrasound. Three days later, the tumors were scanned with a L-14/5 38 array transducer connected to a SonixOne scanner with and without a slab of pork belly placed on top of the tumors. The pork belly acted as an additional source of attenuation and transmission loss. QUS parameters, specifically effective scatterer diameter (ESD) and effective acoustic concentration (EAC), were calculated using calibration spectra from both an external reference phantom and the Ti bead.
Results:
For ESD estimation, the 95% confidence interval between measurements with and without the pork belly layer was (6.0,27.4) using the in situ bead and (114, 135.1) with the external reference phantom. For EAC estimation, the 95% confidence interval were (-8.1, 0.5) for the bead and (-41.5, -32.2) for the phantom. These results indicate that the in situ bead method shows reduced bias in QUS estimates due to intervening tissue losses.
Conclusions:
The use of an in situ Ti bead as a radiological marker not only serves its traditional role but also effectively acts as a calibration target for QUS methods. This approach accounts for attenuation and transmission losses in tissue, resulting in more accurate QUS estimates and offering a promising method for enhanced disease state classification in clinical settings.
More Related Videos
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
09:57Characterizing Microbiome Dynamics – Flow Cytometry Based Workflows from Pure Cultures to Natural Communities
Published on: July 12, 2018
Related Concept Videos
Calibration Curves: Correlation Coefficient
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...