Distances and angles in standing long-leg radiographs: comparing conventional radiography, digital radiography, and
Christof Birkenmaier1,2, Louise Levrard3, Carolin Melcher4,5
1Department of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU Munich, Großhadern Campus, Marchioninistr. 15, 81377, Munich, Germany. christof.birkenmaier@med.uni-muenchen.de.
Objective:
Distances and angles measured from long-leg radiographs (LLR) are important for surgical decision-making. However, projectional radiography suffers from distortion, potentially generating differences between measurement and true anatomical dimension. These phenomena are not uniform between conventional radiography (CR) digital radiography (DR) and fan-beam technology (EOS). We aimed to identify differences between these modalities in an experimental setup.
Materials And Methods:
A hemiskeleton was stabilized using an external fixator in neutral, valgus and varus knee alignment. Ten images were acquired for each alignment and each modality: one CR setup, two different DR systems, and an EOS. A total of 1680 measurements were acquired and analyzed.
Results:
We observed great differences for dimensions and angles between the 4 modalities. Femoral head diameter measurements varied in the range of > 5 mm depending on the modality, with EOS being the closest to the true anatomical dimension. With functional leg length, a difference of 8.7% was observed between CR and EOS and with the EOS system being precise in the vertical dimension on physical-technical grounds, this demonstrates significant projectional magnification with CR-LLR. The horizontal distance between the medial malleoli varied by 20 mm between CR and DR, equating to 21% of the mean.
Conclusions:
Projectional distortion resulting in variations approaching 21% of the mean indicate, that our confidence on measurements from standing LLR may not be justified. It appears likely that among the tested equipment, EOS-generated images are closest to the true anatomical situation most of the time.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Electronic Distance Measuring Instruments