Standardized Data Acquisition for Neuromelanin-Sensitive Magnetic Resonance Imaging of the Substantia Nigra
Published on: September 8, 2021
Comparison of 3D Magnetization-transfer- and Spectral-presaturation-with-inversion-recovery-based Neuromelanin
Midori Kusama1, Yukio Kimura1, Masami Yoneyama2
1Department of Radiology, National Center of Neurology and Psychiatry, Kodaira, Tokyo, Japan.
Purpose:
Neuromelanin is visualized by optimizing the conditions of longitudinal relaxation (T1)-weighted imaging (T1WI). Although it was originally developed in 2D imaging, 3D imaging has been also reported, and T1WI sequences with magnetization transfer (MT) pulses are now widely used in 3D gradient echo (GRE) sequences. In this study, we assert that the use of spectral presaturation with inversion recovery (SPIR) may also be useful as an alternative to MT pulses, and we optimize SPIR and compare it with MT.
Methods:
Neuromelanin images with MT pulse and SPIR (flip angles [FAs] = 19º, 22º, and 25º) were acquired from 30 healthy volunteers. To achieve the same acquisition time of 5 min, the slab thickness of the MT images was less than 1/3 of those of the SPIR images; the acquisition areas for MT and SPIR were the brainstem and the whole brain, respectively. Visual and quantitative evaluation was performed and compared on the four sequences acquired for the substantia nigra pars compacta (SNc) and the locus coeruleus (LC). For visual assessment, we used the mean score from a 3-point scale by two evaluators. For quantitative evaluation, the contrast ratios of SNc and LC were calculated in comparison with the background tissue signal.
Results:
In visual assessments, the mean scores of the SPIR FA19º and FA22º images were better than others in the SNc. Regarding LC, the SPIR FA22º image yielded the best mean score. In quantitative evaluations, the MT image was significantly lower than the other three images in SNc. Regarding LC, there were no significant differences among the four acquired images (MT and SPIR FA19º, FA22º, and FA25º).
Conclusions:
Detection of neuromelanin in SNc and LC was improved by the use of SPIR compared to MT pulse in 3D neuromelanin imaging.
More Related Videos
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
11:38In vivo Imaging of Optic Nerve Fiber Integrity by Contrast-Enhanced MRI in Mice
Published on: July 22, 2014
Related Concept Videos
Magnetic Resonance Imaging
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)