[Research on the Quality Control of Routine Reusable Pipeline and Disposable Pipeline of Ventilator]
Li Bao1, Yunming Shen1, Siwei Xiang1
1Children's Hospital, Zhejiang University School of Medicine, Hangzhou, 310057.
Objective:
To explore the effect of routine reusable pipeline and disposable pipeline on ventilator quality control results.
Methods:
17 ventilators were randomly selected to conduct quality control using routine reusable pipeline and disposable pipeline respectively. Quality control data were recorded and then paired t-test method was used to analyze whether the difference between the two pipelines was significant or not.
Results:
There were no significant differences in respiratory rate, tidal volume and end airway pressure between the two types of pipes ( P>0.05). The airway peak pressure of routine reusable pipeline was significantly higher than disposable pipeline ( P<0.05), but the difference was very small, only about 0.2 mbar which would not affect the conclusion of quality control.
Conclusion:
Quality control of ventilator is not affected by routine reusable pipeline and disposable pipeline, which can be replaced by each other.
More Related Videos
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...