Approaches for Performance Verification Toward Standardization of Peripheral Blood Regulatory T-Cell Detection by
Mei Liu1,2, Jin-Peng Liu1,2, Pan Wang1,2
1From NHC Key Laboratory of AIDS Immunology - China Medical University (M Liu, J-P Liu, Wang, Fu, Jiang, Zhang, Shang), National Clinical Research Center for Laboratory Medicine (M Liu, J-P Liu, Wang, Fu, Zhao, Jiang, Zhang, Shang), and the Department of Laboratory Medicine (Zhao, Shang), The First Hospital of China Medical University, Shenyang, China.
This study validates flow cytometry assays for regulatory T-cell (Treg) detection in peripheral blood. It confirms reliable performance across different reagent sets, aiding standardization for immune-mediated disease diagnosis.
Area of Science:
- Immunology
- Clinical Diagnostics
- Flow Cytometry
Background:
- Regulatory T-cell (Treg) detection via flow cytometry is crucial for diagnosing and managing immune-mediated diseases.
- Standardization and reliable performance verification of Treg assays are currently lacking.
Purpose of the Study:
- To standardize Treg assays by evaluating the performance of three commercial reagent sets.
- To assess Treg detection agreement across different reagent sets using flow cytometry.
Main Methods:
- Analytical performance of Treg assays was evaluated, including gating strategy, antibody optimization, sample stability, repeatability, reproducibility, reportable range, linearity, and assay carryover.
- Agreement between assays was analyzed using Bland-Altman plots and linear regression.
- The correlation between CD4+CD25+CD127low/- Tregs and CD4+CD25+Foxp3+ Tregs was assessed.
Main Results:
- Postcollection sample stability was confirmed up to 72 hours at room temperature.
- Assay accuracy, repeatability, and reproducibility met clinical analysis requirements, with excellent linearity (R2 ≥0.9) and no carryover.
- Results from three manufacturers showed good agreement, and CD4+CD25+CD127low/- Treg frequency correlated well with CD4+CD25+Foxp3+ Tregs.
Conclusions:
- This study provides the first systematic evaluation of Treg measurement performance in peripheral blood using flow cytometry.
- The findings offer a practical method for verifying the performance of flow cytometry-based immune monitoring assays in clinical settings.
- The validated methods contribute to the standardization of Treg detection for improved clinical practice.