Radiation protection in CT-guided interventions: does real-time dose visualisation lead to a reduction in radiation
A Koch1, T Gruber-Rouh1, S Zangos1
1Department of Diagnostic and Interventional Radiology, Frankfurt-University Hospital, Theodor-Stern Kai 7, 60590 Frankfurt am Main, Germany.
Aim:
To evaluate if real-time dose visualisation during computed tomography (CT)-guided interventions leads to a reduction in radiation dose to participating radiologists.
Materials And Methods:
The individual radiation dose radiologists are exposed to during CT interventions was measured using dedicated dosimeters (RaySafe i2-system, Unfors RaySafe GmbH, Billdal, Sweden) worn over the usual radiation protective apron. Initially, only the total radiation dose was measured, without visualisation (control group). In the following study period, the radiation dose was shown to participants on a live screen in real-time (experimental group). In both groups, the dose was recorded in 1-second intervals. The results collected were evaluated by comparison using descriptive statistics and mixed-effect models. In particular, the variables experience, gender, role, and position during the intervention were analysed.
Results:
In total, 517 measurements of 304 interventions (n=249 with and n=268 without live screen) performed by 29 radiologists acting as interventionalists or assistants were analysed. All CT-guided interventions were performed percutaneously, the majority of which (n=280) were microwave ablations (MWA). Radiation doses in the group without visualisation were comparable with usual dose rates for the corresponding intervention type. The mean total radiation dose was reduced by 58.1% (11.6 versus 4.86 μSv) in the experimental group (p=0.034). The highest reduction of 78.5% (15.55 versus 3.35 μSv) was observed in radiologists with the role of assistant (p=0.002). Sub-analysis showed significant dose reduction (p<0.0001) for the use of live screen in general; considering all variables, the role "assistant" alone had a statistically significant influence (p=0.002).
Conclusion:
The real-time visualisation of active radiation dose during CT interventions leads to a relevant reduction in radiation dose to participating radiologists.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...