Analysis of cyclic fatigue resistance of different endodontic nickel-titanium rotary instruments: An in vitro study
Akshayraj Langaliya1, Neeta Patel1, Athulya Pallipurath1
1Department of Conservative Dentistry and Endodontics, AMC Dental College and Hospital, Ahmedabad, Gujarat, India.
Aims:
The aim of this in vitro study was to compare the cyclic fatigue resistance of three different endodontic nickel-titanium rotary instruments using a dynamic testing device.
Materials And Methods:
Ten files each of ProTaper Gold (PG), Hyflex Electro-discharge Machining (HEDM), and TruNatomy (TN) were tested in a custom-fabricated dynamic cyclic fatigue testing device at 60° curvature having a radius of curvature of 5 mm. The number of cycles to the fracture (NCF) of each instrument was calculated and three continuous groups were compared by the Kruskal-Wallis test and Dunn post hoc test was used for pairwise comparison.
Results:
Cyclic fatigue resistance of HEDM was the highest, followed by TN. PG had the lowest among the three.
Conclusion:
Within the limitations of the present in vitro results, it can be concluded that HEDM files appeared to be suitable for shaping complex canals with the greater number of cycles before it fractures.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
08:47Mimicking and Measuring Occlusal Erosive Tooth Wear with the "Rub&Roll" and Non-contact Profilometry
Published on: February 2, 2018
Related Concept Videos
Fatigue
Residual Stresses in Circular Shafts
Stress Concentrations in Circular Shafts
Thin-Walled Hollow Shafts