Related Experiment Videos
Pyrimidine nucleoside phosphorylase assay by automated high-performance liquid chromatography
Abstract:
A new assay for pyrimidine nucleoside phosphorylase is reported. This method utilizes an isocratic reversed-phase high-performance liquid chromatographic system for separation of nucleosides and bases. Product detection is accompanied by ultraviolet monitoring and radioactive flow detection. Use of an automated sample injector allows for the analysis of a series of samples, with data recorded onto a microprocessor-based cassette recorder. Data can then be downloaded into computer memory. The velocity of uridine phosphorylase (E.C. 2.4.2.3) was a linear function of enzyme concentration. The Michaelis constant for uridine at pH 8.0 was found to be in close agreement with the value obtained by a thin-layer chromatographic assay method.